• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Catalog Number 0.850.0009
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 07/05/2016
Event Type  malfunction  
Manufacturer Narrative
The analysis of the returned device has been completed.Our records indicate that the device was assembled in (b)(6) 2010.The returned unit was inspected visually and by x-ray.The investigation found evidence of heavily applied dental office cleaners and fluid ingress into the interior of the handset.It was determined that there was a short circuit in the battery pack circuit board that lead to the battery failure.No injuries were reported.The nomad operator manual provides instructions on the proper care, cleaning, and maintenance of the device.This concludes our investigation.The company initially determined that this event was not reportable, but aribex initiated a recall, reference recall z2716/17-2016.
 
Event Description
Customer reported that battery started to smoke while in use and was very hot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6046161
MDR Text Key58018949
Report Number1017522-2016-00026
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 10/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0.850.0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2016
Initial Date FDA Received10/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/02/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-