The analysis of the returned device has been completed.Our records indicate that the device was assembled in (b)(6) 2010.The returned unit was inspected visually and by x-ray.The investigation found evidence of heavily applied dental office cleaners and fluid ingress into the interior of the handset.It was determined that there was a short circuit in the battery pack circuit board that lead to the battery failure.No injuries were reported.The nomad operator manual provides instructions on the proper care, cleaning, and maintenance of the device.This concludes our investigation.The company initially determined that this event was not reportable, but aribex initiated a recall, reference recall z2716/17-2016.
|