MAUDE Adverse Event Report: HILL-ROM BATESVILLE ADVANTA FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE ADVANTA FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1600B005348
Device Problems Device Alarm System (1012); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2016
Event Type  malfunction  
Manufacturer Narrative
Hill-rom technical support suggested replacing the side communication board.Per the hill-rom service manual the advanta bed requires an effective maintenance program.Preventative maintenance will minimize downtime due to excessive wear.Communications: inspect and test the communication junction box.Test the sidecom communication system features for proper operation.Inspect the communication cable including the male and female pins in the plug.Test the nurse call super capacitor for proper operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
 
Event Description
Hill-rom received a report from the account stating the bed exit alarm was not calling the nurses station.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key6046605
MDR Text Key58356308
Report Number1824206-2016-00457
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP1600B005348
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-