Model Number 1029754
Device Problems Failure To Adhere Or Bond (1031); Material Fragmentation (1261)
Patient Problem Missing Value Reason (3192)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.
 
Event Description
Country of complaint: (b)(4).An issue was reported about the bone wax.The product crumbles, does not stick well to the instruments and is not effective.
 
Manufacturer Narrative
Samples received: 4 unopened units (each one of one different batch).Analysis and results: there are no previous complaints of any of the involved code-batches.(b)(4).There are no units in stock.Batch manufacturing records of the involved reference-batches have been checked and no deviations have been found.The following parameters of each sample received and other three different batches have been tested: smell, color, appearance, suppleness and adhesive behavior.All batches are into the current range for this product.No differences have been found between the different batches analyzed: - the smell is the usual one, there is not smell rancid.- the color is white yellow as usual.- the appearance is according to the visual standard.- regarding the suppleness, the products can be kneaded well.- and concerning the adhesive behavior, the product hold well on instruments.Final conclusion: although the results of the samples received fulfill the oem specifications, note of this incident is taken in order to assess if new or additional actions are needed.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE WAX 24X2.5G
Type of Device
BONE WAX
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6048449
MDR Text Key58506161
Report Number2916714-2016-00877
Device Sequence Number1
Product Code MTJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K000021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1029754
Device Catalogue Number1029754
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Distributor Facility Aware Date10/06/2016
Initial Date Manufacturer Received 09/22/2016
Initial Date FDA Received10/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/31/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-