Model Number AG-920RA
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/23/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reports that the multi gas unit will lose connection with the main monitor and the readings blank out.The device also starts reading carbon dioxide when it isn't sampling air, and it will just continue to cycle through looking for other gases when it shouldn't.Customer would like to send device in for repair.The unit was cleaned and evaluated.The reported problem of unit loosing connection with main monitor was not duplicated.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit was tested and operates to manufacturer's specifications.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reports that the multi gas unit will lose connection with the main monitor and the readings blank out.The device also starts reading carbon dioxide when it isn't sampling air, and it will just continue to cycle through looking for other gases when it shouldn't.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn; susan shadley
tomioka city, japan 370-2 314
JA  370-2314
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: susan shadley
tomioka city, japan 370-2 314
JA   370-2314
Manufacturer Contact
susan shadley
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8-560
JA   161-8560
2687133
MDR Report Key6049481
MDR Text Key58130476
Report Number8030229-2016-00531
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/21/2016,09/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/21/2016
Distributor Facility Aware Date09/23/2016
Device Age96 MO
Event Location Hospital
Date Report to Manufacturer10/21/2016
Date Manufacturer Received10/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-