Catalog Number RF410 SMN10444910 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted the siemens customer care center (ccc).Siemens customer support determined that the aeration ball on the water filtration system was not working properly.The customer contacted the water filtration vendor and replaced the aeration ball.The cause for the discordant falsely low ft4 result is unknown.The device is performing within specifications.No further evaluation of the device is required.
|
|
Event Description
|
A discrepant low free thyroxine (ft4) result was obtained on a patient sample on the dimension xpand system.The patient result was reported to the physician who questioned the result.The same sample was sent for testing at an alternate laboratory and run using a different methodology.A higher result was obtained.The physician regarded the higher result as correct.There is no indication that patient treatment was altered or prescribed on the basis of the discrepant low ft4 result.There was no report of adverse health consequences as a result of the discrepant low ft4 result.
|
|
Search Alerts/Recalls
|