• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0009
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 09/27/2016
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to the manufacturer and an evaluation is pending.A follow-up report will be filed upon investigation completion.A recall was initiated, reference (b)(4).
 
Event Description
It was reported that the battery was placed on the charger overnight and it melted.
 
Manufacturer Narrative
Investigation pending.A follow-up report will be submitted upon investigation completion.
 
Manufacturer Narrative
Upon visual inspection, it appears that a thermal runaway occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and pcb.There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes our investigation.A recall is ongoing, reference z-2716/2717-2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6058940
MDR Text Key58454815
Report Number1017522-2016-00034
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date 09/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Device Catalogue Number0.850.0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-