• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0009
Device Problems Thermal Decomposition of Device (1071); Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2016
Event Type  malfunction  
Manufacturer Narrative
Unit has not yet been returned for investigation.A follow-up report will be submitted upon investigation.A recall has been initiated, reference (b)(4).
 
Event Description
It was reported that there was a burning smell coming from the unit.
 
Manufacturer Narrative
Unit received by manufacturer on 11/28/2016 and will undergo investigation.A follow-up report will be submitted upon investigation completion.
 
Manufacturer Narrative
There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal runaway, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and pcb.This concludes our investigation.A recall is ongoing, reference z-2716/2717-2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6059178
MDR Text Key58845337
Report Number1017522-2016-00035
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Device Catalogue Number0.850.0009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/30/2016
Initial Date FDA Received10/26/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/01/2016
02/23/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/13/2011
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-