• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3721
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 10/04/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).This charger model was included in a field correction.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating udi (di): sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
Device #2 of 2: reference mfr.Report: 1627487-2016-05605.It was reported the patient is experiencing warmth at the ipg site after charging.Surgical intervention may be pending to address this issue.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device #2 of 2: reference mfr.Report: 1627487-2016-05604.Follow up information identified the patient underwent surgical intervention on (b)(6) 2016 where the ipg was removed and replaced.Issue has been resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6062151
MDR Text Key58599554
Report Number1627487-2016-05605
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2010
Device Model Number3721
Device Lot Number2889953
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
MODEL 3219, SCS LEAD
Patient Outcome(s) Other;
Patient Age68 YR
-
-