• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem Not Applicable (3189)
Event Date 10/06/2016
Event Type  malfunction  
Manufacturer Narrative
On 10/06/2016 the field service engineer (fse) found a leak in tubing through fitting ff421 which delivers vacuum to clean the probe after aspiration.The fse replaced the tubing and the instrument ran without leaks.The repairs were verified per established procedures.(b)(4).
 
Event Description
The customer reported a contained diluent/blood leak of less than 10 ml from the unicel dxh 800 coulter cellular analysis system.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results and controls.The customer was wearing personal protective equipment (ppe) consisting of laboratory coat and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key6063861
MDR Text Key58666885
Report Number1061932-2016-00881
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number629029
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-