Model Number 97714
Device Problems Failure to Deliver Energy (1211); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problems Pain (1994); Therapeutic Effects, Unexpected (2099); Therapeutic Response, Decreased (2271)
Event Date 09/28/2016
Event Type  malfunction  
Manufacturer Narrative
Product id: 37761, serial# (b)(4), product type: recharger.
 
Event Description
A consumer implanted for failed back surgery syndrome and spinal pain reported on (b)(6) 2016 the connector pin was bent, so the recharger stopped charging.It was further reported on (b)(6) 2016 stimulation stopped working there was sudden return of pain.No out of box failure was reported.
 
Manufacturer Narrative
Analysis of the desktop charger with sn# (b)(4) found that the desktop charger had broken connector tip.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received confirmed that the desktop charger had a broken connector pin.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6064016
MDR Text Key58721554
Report Number3004209178-2016-22790
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2016
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/29/2016
Initial Date FDA Received10/28/2016
Supplement Dates Manufacturer ReceivedNot provided
03/07/2017
Supplement Dates FDA Received03/07/2017
09/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
-
-