Model Number 97714 |
Device Problems
Failure to Deliver Energy (1211); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problems
Pain (1994); Therapeutic Effects, Unexpected (2099); Therapeutic Response, Decreased (2271)
|
Event Date 09/28/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product id: 37761, serial# (b)(4), product type: recharger.
|
|
Event Description
|
A consumer implanted for failed back surgery syndrome and spinal pain reported on (b)(6) 2016 the connector pin was bent, so the recharger stopped charging.It was further reported on (b)(6) 2016 stimulation stopped working there was sudden return of pain.No out of box failure was reported.
|
|
Manufacturer Narrative
|
Analysis of the desktop charger with sn# (b)(4) found that the desktop charger had broken connector tip.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received confirmed that the desktop charger had a broken connector pin.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|