It was reported that the egr system blade is not retracting.The device was not in contact with the patient, no patient injury reported however, revision/medical intervention was required.Additional information was requested, no further information will be provided.
|
Integra has completed their internal investigation on november 7, 2016.The investigation included: methods: review of device history records review of complaints history results: dhr review; egr 157 assembly lot # pa0022, consisting of (b)(4) units, was assembled on april 22 ¿ 28, 2015 and inspected on april 24 ¿ 29, 2015.Final audit activities at pti involved a 100% inspection of each egr assembly.Packaged and sterilized egr 157 lot # pa0022a1, consisting of (b)(4) pieces, was received at (b)(4) on june 19, 2015, inspected (n = 8) and released into finished goods inventory on june 22, 2016.There were no material non-conformances or variance associated with inspection of the final product.Complaints history; a query in our complaint management system determined that fourteen (14) other customer complaints have been initiated based on the inability for the egr cutting blade to retract.The query was based on the individual assembly part numbers 31-0001, 31-0101, the system catalog numbers 310040 and 310138 and the terms ¿endoscopic¿ and ¿egr¿ and covered the time frame of january 1, 2014 ¿ october 25, 2016.Conclusion: the egr 157 device associated with this complaint was not returned for evaluation.As a result, a likely root cause can be identified since integra is unable to determine if the reported non-conformance was due to an out-of-box failure, mechanical failure during use or from user technique.Integra lifesciences continues to monitor the field for all issues associated with the egr devices, and further action will be taken if any new risks or adverse trends are identified.
|