MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-523LNAS
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Naturally Worn (2988)
Patient Problems Hyperglycemia (1905); Diabetic Ketoacidosis (2364)
Event Date 10/26/2016
Event Type  Injury  
Manufacturer Narrative
Currently, it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.(b)(4).
 
Event Description
The customer reported via phone call of being hospitalized for high blood glucose of 565mg/dl and diabetic ketoacidosis.Troubleshooting found that the keypad buttons are not responsive and the arrow buttons are worn down to the metal.Customer was advised to discontinue use of the device and revert to a back-up plan per their doctor's instruction.The customer was also advised that the device would be replaced and agreed to return the product for analysis.Customer declined high blood glucose troubleshooting.
 
Manufacturer Narrative
The insulin pump passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.However, insulin pump was received with intermittent button response.The insulin pump was received with cracked battery tube threads, keypad overlay worn and minor scratched lcd window.
 
Manufacturer Narrative
The insulin pump passed the displacement accuracy test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6073460
MDR Text Key59008005
Report Number2032227-2016-39397
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 04/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/28/2016
Initial Date FDA Received11/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/01/2016
04/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age43 YR
Patient Weight59
-
-