Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2016
Event Type  malfunction  
Event Description
It was reported that during a cryoablation procedure, shortly after entering the balloon catheter into the sheath, an air bubble was noticed in the sheath side port.A gentle aspiration of the sheath removed the bubble.The physician noted that it was possible the bubble was introduced by a defective sheath side port.The sheath was replaced and the case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Event summary: the clinical data files showed at least fifteen injections were performed without any issue on the date of the event.Upon visual inspection of flexcath sheath (b)(4), results show the device was intact with no apparent issues.Dissection showed that the hemostatic valve was leaking and the valve was torn.In conclusion, the reported issue has been confirmed through testing.The sheath failed the returned product inspection due to a leaking hemostatic valve.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6074011
MDR Text Key59018550
Report Number3002648230-2016-00472
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date08/05/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number58736
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial