Siemens became aware of the reported issue on october 3, 2016 and the investigation in on-going.A supplemental report will be submitted upon completion.This information has been requested and has not been received, to date.Should the information become available a supplemental report will be submitted at the required regulatory timelines.
|
The customer informed siemens on october 3, 2016 that during manual positioning of a patient in the treatment room, either the table locked by itself or it changed vertical high speed to low by itself.Reportedly, on (b)(6) 2016, the patient was manually positioned on the table.The gantry position was set at the treatment console by pushing f12, alt enable and the gantry moved to position.However, the table made a longitudinal movement of 20 cm from the gantry at the same time of the gantry movement.It was also reported that on (b)(6) 2016, the patient was manually positioned on the table.The gantry position was set at the treatment console by pushing f12, alt enable to move to position.The table position was also set at the treatment console by pushing f12, alt enable to iso 310' position.However, movement of the gantry and table occurred at the same time with the table making a longitudinal movement of 8.5 cm from the gantry.There are no reports of injury or mistreatment to a patient.This reported issue occurred in (b)(6).
|