• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON 780-15 CIRCUIT W/ COLUMN; HUMIDIFIER NEBULIZER KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON 780-15 CIRCUIT W/ COLUMN; HUMIDIFIER NEBULIZER KIT Back to Search Results
Catalog Number 780-15KIT
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 10/27/2016
Event Type  malfunction  
Manufacturer Narrative
Qn#(b)(4).Conclusion: the complaint was confirmed, but the root cause could not be identified.A video was received as part of the complaint information.The video shows a circuit under water and it can be observed bubbles between the jar & water trap that appears to have a leak between both components.The device history record of batch number 74a1602675 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Customer complaint is confirmed based on video provided that shows a circuit that appears to have a leak between the jar and the water trap.However, physical sample is necessary to conduct a proper investigation and determine the source of defect reported.If device sample becomes available at a later date this complaint will be updated.
 
Event Description
The customer alleges that the infant circuit failed the pre-test.The customer alleges that there is a leak at the location where the cup screws into the housing (trap).No patient injury reported.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and no tube melting or charring was detected.The entire circuit was leak tested.In order to thoroughly capture all leaks, if any, both limbs of the circuit were submerged under a bath of plain tap water to see if any bubbles could be detected from leaks.Immediately leaks were detected at both rain trap cups, one on the inspiratory limb, the other on the expiratory limb.The leaks were coming from where the collection cup attaches by plastic threads to the rain trap base.These leaks were so profuse an acceptable rate of leak volume was unobtainable.Based on the investigation performed, the reported complaint was confirmed.The returned circuit was confirmed to leak from the expiratory and inspiratory limb water traps.A dhr review was performed on the product with no evidence to suggest a manufacturing related cause.All circuits are 100% inspected for leaks at the time of manufacturing; therefore, any defects would be detected prior to release.This defect was discovered during preliminary testing (prior to use), sst testing on a ventilator.Other remarks: based upon the time of discovery and the damage observed on the returned sample, it was determined that operational context caused or contributed to this event.
 
Event Description
The customer alleges that the infant circuit failed the pre-test.The customer alleges that there is a leak at the location where the cup screws into the housing (trap).No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON 780-15 CIRCUIT W/ COLUMN
Type of Device
HUMIDIFIER NEBULIZER KIT
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6076891
MDR Text Key59083905
Report Number3004365956-2016-00411
Device Sequence Number1
Product Code OGG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date10/07/2020
Device Catalogue Number780-15KIT
Device Lot Number74A1602675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2016
Initial Date FDA Received11/03/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SERVO-I VENTILATOR
-
-