Device Problem Fluid/Blood Leak (1250)
Patient Problems Pain (1994); Weight Changes (2607)
Event Date 09/19/2016
Event Type  malfunction  
Manufacturer Narrative
Unknown taper.Medwatch submitted to the fda on 11/03/2016.The reporter of the event was asked to return the product for analysis, and to indicate product serial number.To date, neither the device nor any further device information has been received by apollo.Without device or device serial, the taper type is unknown.If returned, visual examination may determine the connector type associated with this event.This event was reported by the patient.Apollo has been able to confirm the reported events of leak and pain with the patient's physician.This submission represents the patients second port leak.The patient has had a previous leak which will be captured under mdr submission mdr 3006722112 2016 00328.Device labeling to address the reported events: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Other adverse events considered related to the lap-band system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warning: patients should be advised that the lap-band ap system is a long-term implant.Explant (removal)and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Reported as: a patient with the lap-band system was reported to have "experiencing "ulcer" like pain in the side of the abdomen.Patient believes the port is "broke again." patient does not feel good, has stomach aches, can feel something jagged in the side of the abdomen, and has regained weight." the device currently remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM ACCESS PORT
Type of Device
ADJUSTABLE GASTRIC BAND ACCESS PORT
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capitol of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia, costa rica
CS  
Manufacturer Contact
laura leboeuf
1120 s capitol of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key6077727
MDR Text Key59128574
Report Number3006722112-2016-00306
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Patient
Type of Report Initial
Report Date 10/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/05/2016
Initial Date FDA Received11/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age44 YR
Patient Weight88
-
-