On (b)(6) 2016, the reporter contacted animas, alleging a history/settings (inaccurate delivery) issue.No details were provided.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported as there is an allegation against the delivery function of the pump.
|
The device has been returned and evaluated by product analysis on 12/19/2016 with the following findings: a review of the pump histories indicated that the last bolus and last basal deliveries occurred on (b)(6) 2016.A review of the black box indicated the following: an ¿exceeded max 2 hour limit¿ warning occurred on (b)(6) 2016 at 19:24 and deliveries resumed at 20:03; an unexplained loss of prime warning occurred on (b)(6) 2016 at 09:04, the pump was powered off, and deliveries resumed on (b)(6) 2016 at 15:57.A review of the total daily dose history indicated that insulin delivery totals correctly reflected programmed values.The pump successfully completed a rewind, load, and prime sequence.The pump was exercised for 24 hours with no issues occurring.The pump passed delivery accuracy testing and was found to be delivering within required specifications.No delivery interruptions and no errors, alarms, or warnings occurred during investigation.(b)(4).
|