Model Number 3C98-3=9.2
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/27/2015
Event Type  malfunction  
Manufacturer Narrative
The device was incorrectly assembled by the manufacturing department: a missing screw lead to a loose servo-electronic unit; no fall or critical situation occurred.This malfunction of the device could cause or contribute to further falls and potential serious injuries might be possible if the malfunction were to recur.
 
Event Description
Thirty beeps error.Having issues with charging, making beeping noises and vibrations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C-LEG
Type of Device
EXTERNAL ABOVE KNEE PROSTHESIS
Manufacturer (Section D)
OTTO BOCK HEALTHCARE PRODUCTS GMBH
brehmstrasse 16
vienna, vienna 1110
AU  1110
Manufacturer (Section G)
OTTO BOCK HEALTHCARE PRODUCTS GMBH
brehmstrasse 16
vienna, vienna 1110
AU   1110
Manufacturer Contact
reinhard wolkerstorfer
brehmstrasse 16
vienna, vienna 1110
AU   1110
1 5233786
MDR Report Key6082044
MDR Text Key59349194
Report Number9615892-2016-00013
Device Sequence Number1
Product Code ISY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number3C98-3=9.2
Device Catalogue Number3C98-3=9.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-