• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM FRANCE CAREASSIST ES110/210/410, NUL; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM FRANCE CAREASSIST ES110/210/410, NUL; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1170D0000041
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2016
Event Type  malfunction  
Manufacturer Narrative
Hill-rom technical support suggested calibrating the scale.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Every five years, the bed exit tape switches should be replaced by a facility-authorized maintenance personnel.Ensure the bed exit system works properly.Replace worn or defective parts as needed.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Three attempts have been made regarding a resolution to this contact line, with no response.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed exit was not alarming.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST ES110/210/410, NUL
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM FRANCE
188 rue du caducee-parc , euromedecine
montpellier, herault 34195
FR  34195
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key6083196
MDR Text Key59630079
Report Number9615739-2016-00084
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP1170D0000041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-