A field service engineer (fse) evaluated the analyzer on 10/14/2016 and confirmed a leak at the end of the probe, caused by a blockage at the probe wipe rise block.Additionally, and unrelated to the leak, the hemoglobin lamp was worn and the analyzer alerted to hemoglobin voltage high errors.The fse replaced the probe wipe rinse block to correct the probe leaking issue, and also replaced and adjusted the hemoglobin lamp to service specifications.
|
A customer reported a leak of less than 2 ml of fluid from the probe of a coulter ac·t diff 2 analyzer; additionally, hemoglobin voltage high errors were recovered.The operator was wearing personal protective equipment (ppe), consisting of gloves, goggles and a laboratory coat at the time the leak was observed.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
|