Model Number MMT-1711B |
Device Problems
Difficult or Delayed Positioning (1157); Naturally Worn (2988); Physical Property Issue (3008)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 10/12/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.(b)(4).
|
|
Event Description
|
It was reported of damage from the insulin pump.The customer's blood glucose reading was 16.8 mmol/l.The customer reported that the serial number has worn off the back of the pump.The customer also stated that the down arrow has to be pushed very hard sometimes.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
The device had fading serial number label, scratched case and a pillowing keypad overlay.All buttons functioned properly.No damage noted on the keypad traces.During visual inspection, found the keypad connector was properly locked.
|
|