MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-1711B
Device Problems Difficult or Delayed Positioning (1157); Naturally Worn (2988); Physical Property Issue (3008)
Patient Problem Hyperglycemia (1905)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.(b)(4).
 
Event Description
It was reported of damage from the insulin pump.The customer's blood glucose reading was 16.8 mmol/l.The customer reported that the serial number has worn off the back of the pump.The customer also stated that the down arrow has to be pushed very hard sometimes.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
The device had fading serial number label, scratched case and a pillowing keypad overlay.All buttons functioned properly.No damage noted on the keypad traces.During visual inspection, found the keypad connector was properly locked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6086647
MDR Text Key59748140
Report Number2032227-2016-40480
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1711B
Device Catalogue NumberMMT-1711B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/12/2016
Initial Date FDA Received11/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-