MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2016
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturing representative (rep).Telemetry problem with 8840 was reported; the rep was in a case and the pump was out of the box on the back table, they were initially able to interrogate the pump and at the time of the call, they could not establish telemetry.There were no symptoms reported.The rep was going to try turning off the lights to see if that would help resolve the issue and call back if they could not resolve the issue.It was discussed that they can use the backup pump if they cannot establish telemetry on this pump again (pump was already in sterile field on back table and could not be moved).Additional information received from the manufacturer representative.It was reported that the pump was never implanted.They did not test again to see if telemetry could be established.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The company representative had the pump and was to return it.
 
Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.
 
Manufacturer Narrative
The pump (b)(4) was returned for analysis.Upon interrogation the pump was used to deliver water (1000 mcg/ml at 6.4 mcg/day).Analysis of the pump found no anomalies.The conclusion code (b)(4) no longer applies.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6087778
MDR Text Key59459386
Report Number3004209178-2016-23427
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/14/2018
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2016
Initial Date FDA Received11/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
01/10/2017
Supplement Dates FDA Received11/16/2016
12/08/2016
01/17/2017
09/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-