• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-554LCAH
Device Problems Device Alarm System (1012); Bent (1059); No Flow (2991)
Patient Problem Hyperglycemia (1905)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.(b)(4).
 
Event Description
It was reported that the device was not delivering insulin and device did not alarm no delivery alarm.Customer's blood glucose was 12.5 mmo/l.Infusion set cannula was bent.After troubleshooting, customer was advised a tubing clamp will be sent to perform the high pressure test.Customer will not be returning the device for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6091115
MDR Text Key59874692
Report Number2032227-2016-40919
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-554LCAH
Device Catalogue NumberMMT-554LCAH
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/04/2016
Initial Date FDA Received11/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-