U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-722NAB
Device Problems Device Displays Incorrect Message (2591); No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
The insulin pump was received with compromised force sensor system alarm during basic occlusion test due to loose/protruded drive support disk.The insulin pump had cracked case at display window corner, minor scratched lcd window and cracked reservoir tube lip.(b)(4).
 
Event Description
The customer reported via phone call that they received compromised force sensor system alarm.The customer also reported that they were unable to fill the tubing.The customer's blood glucose was unknown at the time of incident.The customer was advised that the insulin pump will need to be replaced.The customer was advised to disconnect from the insulin pump and revert to back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6091136
MDR Text Key59876161
Report Number2032227-2016-40963
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-722NAB
Device Catalogue NumberMMT-722NAB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/17/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age79 YR
Patient Weight107
-
-