ZIMMER, INC. ZIMMER NATURAL NAIL CALIBRATED DRILL; TRAUMA INSTRUMENT Back to Search Results
Catalog Number 00249006450
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This report will be amended when our investigation is complete.
 
Event Description
It is reported the sleeve tip was fractured when it was pulled to check its function due to the nail not being able to be inserted.Fractured pieces were all collected.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Hardness and dimension taken are within spec.Item is fractured into multiple pieces and exhibits wear marks.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER NATURAL NAIL CALIBRATED DRILL
Type of Device
TRAUMA INSTRUMENT
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
kathleen smith
p.o. box 708
warsaw, IN 46580
8006136131
MDR Report Key6095906
MDR Text Key59739825
Report Number0001822565-2016-04161
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00249006450
Device Lot Number62270443
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-