DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE (COLIBRI); INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.001 |
Device Problems
Sticking (1597); Device Inoperable (1663); Defective Component (2292); Incorrect Or Inadequate Test Results (2456); Device Operates Differently Than Expected (2913); Naturally Worn (2988)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: (b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that the small battery drive device had an undetermined malfunction.During the pre-repair diagnostics assessment, it was determined that the control unit was not functioning and was defective.It was further determined that the controller was defective, the lever was stuck,and the motor, bearings and the tool coupling were worn out.It was further determined that the device failed for check the attachment coupling and for check the off/osc/on switch mode function.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|