BIOMET ORTHOPEDICS G7 STRAIGHT INSERTER THREADED SHAFT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 09/29/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.The following sections could not be completed with the limited information provided.
 
Event Description
During a total hip arthroplasty, the acetabular inserter fractured while impacting the liner.No additional information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.It was noted that the fracture pattern is consistent to that of bending overload.Root cause of the reported event was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will contribute to monitor for trends.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STRAIGHT INSERTER THREADED SHAFT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key6096907
MDR Text Key59736015
Report Number0001825034-2016-04677
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110003452
Device Lot Number427424
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-