Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Information (3190)
|
Event Date 09/29/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.The following sections could not be completed with the limited information provided.
|
|
Event Description
|
During a total hip arthroplasty, the acetabular inserter fractured while impacting the liner.No additional information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.It was noted that the fracture pattern is consistent to that of bending overload.Root cause of the reported event was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will contribute to monitor for trends.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.
|
|