ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY QUADRA CRT-D, US MODEL, DF4COMP; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
|
Back to Search Results |
|
Model Number CD3249-40Q |
Device Problems
Premature Discharge of Battery (1057); Failure to Interrogate (1332); Pacing Problem (1439)
|
Patient Problems
Fainting (1847); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/12/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the patient presented in the emergency room with symptoms of pre-syncope that did not have pacing support.Physician was unable to communicate with the device.When the device was interrogated two days before, device longevity was 2.5 years.Interrogating in another room with another programmer did not resolve the issue.Pre-mature battery depletion was suspected.Deice was explanted and replaced without issues.The patient was stable before, during, and after the procedure and will continue to be monitored going forward.The patient was no longer symptomatic following the procedure.
|
|
Manufacturer Narrative
|
No conclusion code available; the reported field event of an inability to interrogate the device was confirmed in the laboratory and was due to premature battery voltage.Based on all available parameter and usage information, device longevity was found to be below expected limits.Review of the device image also confirmed no pacing during backup vvi mode.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
|
|
Manufacturer Narrative
|
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
|
|