BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2J8002
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an intravia container had a flow issue.The bag does not fit properly in the locked case and disrupts the flow of solution.This occurred during use.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed and the device was found to flow normally with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6100759
MDR Text Key59840243
Report Number1416980-2016-17291
Device Sequence Number1
Product Code KPE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K964853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2J8002
Device Lot NumberDR16F06055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
EPIDURAL NARCOTIC
-
-