• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H00-01
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
 
Event Description
During maintenance on the cell-dyn sapphire the field service technician confirmed a burn mark on the back of the board in between j9 and j10 on the printed wiring part number (b)(4).No injuries occurred.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and instrument service.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The customer submitted the affected pcb assy, psm board, list number 8960291001, to aid this investigation.The inspection of the returned pcb assy, psm board found a burnt +24vdc power trace from pin 9, j9 to pin 9, j10 on the pressure sensor module pcb assembly, which may have overheated or have been caused by an overcurrent of an externally connected defective pump relay.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN SAPPHIRE ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6102637
MDR Text Key59888509
Report Number2919069-2016-02304
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740016616
UDI-Public00380740016616
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K051215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H00-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/26/2016
Initial Date FDA Received11/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-