• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON SOLID CR TIBIAL INSERT TRIAL SZ6 - 11MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON SOLID CR TIBIAL INSERT TRIAL SZ6 - 11MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-T-611B
Device Problems Crack (1135); Fracture (1260); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/18/2016
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The 6 by 11 tibial insert trial cracked when surgeon inserted it by hand pressure.He heard it crack and immediately ceased inserting it.The component was removed in tact with a visible crack.No patient consequences.The final implant insert was implanted successfully.
 
Manufacturer Narrative
An event regarding crack/fracture involving a triathlon trial was reported.The event was confirmed.-device evaluation and results: a crack was observed on the returned device.Material analysis was performed and concluded that "the fracture occurred in fast fracture through the center of the device.The origin was found to be inside the material at the interface between the first and second injection molding shots." -medical records received and evaluation: no medical records or x-rays were made available for evaluation.-device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.-complaint history review: there has been no other event for the lot referenced.Conclusions: material analysis was performed and concluded that "the fracture occurred in fast fracture through the center of the device.The origin was found to be inside the material at the interface between the first and second injection molding shots." a capa investigation was initiated for the reported failure mode.The investigation concluded: the parts are injection molded by (b)(4).Sample parts were pulled from fg.Lack of fusion areas were observed between the first and second shots of the two-shot trials.
 
Event Description
A 6 by 11 tibial insert trial cracked when surgeon inserted it by hand pressure.He heard it crack and immediately ceased inserting it.The component was removed in tact with a visible crack.No patient consequences.The final implant insert was implanted successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON SOLID CR TIBIAL INSERT TRIAL SZ6 - 11MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
rita intorrella
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6103088
MDR Text Key60304612
Report Number0002249697-2016-03595
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5530-T-611B
Device Lot NumberKMAVR3E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
Patient Weight48
-
-