• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; OCCLUDER, TRANSCATHETER SEPTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; OCCLUDER, TRANSCATHETER SEPTAL Back to Search Results
Catalog Number GSX0020A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Headache (1880); Thrombus (2101)
Event Date 10/25/2016
Event Type  Injury  
Manufacturer Narrative
A review of the manufacturing records verified the lot was processed normally and all pre-release specifications were met.(b)(4).
 
Event Description
It was reported a physician implanted a 20mm gore cardioform septal occluder to close a patent foramen ovale on (b)(6) 2015 to treat chronic migraines.The headaches resolved post closure.Recently, the patient presented with headaches and the device was imaged.Echocardiography showed no visible shunting; however, a thrombus formation was seen on the right atrial disc.The patient had not been on antiplatelet therapy since 6 months after the implant and is now being treated with anticoagulants.
 
Manufacturer Narrative
Based on additional information received by gore, fields have been updated to (b)(6) 2016.Echocardiographic images (performed (b)(6) 2016) were received by gore for evaluation.Based on the images provided, thrombus formation on the gore cardioform septal occluder cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
OCCLUDER, TRANSCATHETER SEPTAL
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
marci stewart
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key6103740
MDR Text Key59973666
Report Number2017233-2016-00871
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/23/2017
Device Catalogue NumberGSX0020A
Device Lot Number14197780
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
Patient Weight68
-
-