• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. IMPACTOR-ONLAY INSERT; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. IMPACTOR-ONLAY INSERT; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160405
Device Problems Material Fragmentation (1261); Naturally Worn (2988); Insufficient Information (3190)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Plastic tibial impactor worn.Leaving debris when impacting.
 
Manufacturer Narrative
Device was not returned.An event regarding wear involving an impactor-onlay insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as the device was not returned.Medical records received and evaluation: not performed as medical records were not provided for review.Device history review: a device history review confirmed all devices accepted into finished goods conformed to specification.Complaint history review: no other events were reported for the lot indicated.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the instrument are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available, this investigation will be reopened.Not returned to the manufacturer.
 
Event Description
Plastic tibial impactor worn.Leaving debris when impacting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACTOR-ONLAY INSERT
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
joann lavatelli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6105365
MDR Text Key60551149
Report Number3005738134-2016-00005
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160405
Device Lot Number12250914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/19/2016
Initial Date FDA Received11/16/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-