• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD Back to Search Results
Catalog Number 1235250-28
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 11/02/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.The reported patient effects of thrombosis, as listed in the absorb gt1 instructions for use, is a known adverse event associated with the use of a coronary scaffold in native coronary arteries.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.
 
Event Description
It was reported the patient presented with non-st-elevation myocardial infarction (nstemi).The procedure was to treat a lesion in the proximal left anterior descending coronary artery.The lesion was pre-dilated with an unspecified 2.5mm balloon dilatation catheter.The 2.5x28 mm absorb gt1 bioresorbable vascular scaffold (bvs) was deployed without issue.Post-dilatation was performed with 3.0x15 mm and 2.5x20 mm nc balloons.An optical coherence tomography (oct) catheter was advanced in the patient anatomy confirming the scaffold was fully apposed to the vessel wall at which time thrombus was noted in the bvs.The thrombus was treated by increasing medication.The patient is stable.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6106374
MDR Text Key60019593
Report Number2024168-2016-07998
Device Sequence Number1
Product Code PNY
UDI-Device Identifier08717648212987
UDI-Public(01)08717648212987(17)170711(10)6071261
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 11/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/11/2017
Device Catalogue Number1235250-28
Device Lot Number6071261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient Weight70
-
-