U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK PATELLA RESURFACING ROUND DOME-35MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK PATELLA RESURFACING ROUND DOME-35MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 180320-4
Device Problem Insufficient Information (3190)
Patient Problem Injury (2348)
Event Date 10/31/2016
Event Type  Injury  
Manufacturer Narrative
A supplemental report will be submitted when additional information becomes available.Device not available.
 
Event Description
The surgeon did a revision surgery of a mako pfa to tka.
 
Manufacturer Narrative
An event regarding revision involving a mako patella was reported.The event was not confirmed.Method & results: -device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.-medical records received and evaluation: no medical records or x-rays were made available for evaluation.-device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.-complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
The surgeon did a revision surgery of a mako pfa to tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK PATELLA RESURFACING ROUND DOME-35MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6106777
MDR Text Key60035303
Report Number3005985723-2016-00390
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number180320-4
Device Lot Number12111111-2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/28/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age47 YR
-
-