It was reported that during a laparoscopic gastric bypass procedure, when setting up for the procedure, it was noted prior to opening the device that there was a 1/2 centimeter hole in the tyvek lid of the device.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
|
(b)(4).Batch # n91c1y.The analysis results found that harh45 device was returned inside its package.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; it was noted to have a hole in the tyvek.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|