Catalog Number HARH45
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot/batch.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that during a laparoscopic gastric bypass procedure, when setting up for the procedure, it was noted prior to opening the device that there was a 1/2 centimeter hole in the tyvek lid of the device.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Batch # n91c1y.The analysis results found that harh45 device was returned inside its package.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; it was noted to have a hole in the tyvek.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE + 7
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6112475
MDR Text Key60687679
Report Number3005075853-2016-06602
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue NumberHARH45
Device Lot NumberN91C1Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR & HANDPIECE
-
-