MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number NEU_ENTERRA_INS
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2016
Event Type  Injury  
Manufacturer Narrative
Information references the main component of the system and other applicable components are: product id: 4351, serial# (b)(4), product type: lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The health care provider (hcp) via a manufacturer representative reported that when the lead was passed to the surgeon it was ok on (b)(6) 2016.While scoping after lead insertion and before suturing, the surgeon noted the blue monofilament was broken.The propylene was broken.The lead was replaced and would be sent back for analysis.No symptoms were reported.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Analysis results confirmed that the blue monofilament suture was broken.It was noted that it was overstressed and damaged.The material was split and stretched.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information from the rep reported that the surgeon broke the monofilament.No further complications were reported.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6113185
MDR Text Key60355507
Report Number3007566237-2016-04138
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNEU_ENTERRA_INS
Device Catalogue NumberNEU_ENTERRA_INS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-