Model Number NEU_ENTERRA_INS |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/31/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Information references the main component of the system and other applicable components are: product id: 4351, serial# (b)(4), product type: lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The health care provider (hcp) via a manufacturer representative reported that when the lead was passed to the surgeon it was ok on (b)(6) 2016.While scoping after lead insertion and before suturing, the surgeon noted the blue monofilament was broken.The propylene was broken.The lead was replaced and would be sent back for analysis.No symptoms were reported.
|
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Analysis results confirmed that the blue monofilament suture was broken.It was noted that it was overstressed and damaged.The material was split and stretched.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information from the rep reported that the surgeon broke the monofilament.No further complications were reported.
|
|
Manufacturer Narrative
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|