MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP FT4; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP FT4; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number N/A
Device Problems Calibration Problem (2890); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2016
Event Type  malfunction  
Manufacturer Narrative
The cause for the discordant advia centaur xp ft4 results is unknown.The calibration and qc were valid at the time the samples were run.The calibrator used was lot ca90.Siemens healthcare diagnostics is investigating.The interpretation of results section of the instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings.".
 
Event Description
The customer observed two samples with depressed advia centaur xp ft4 results compared to the clinical picture of the patients.The results were reported to the doctor and the doctor questioned the results.There are no reports that treatment was altered, prescribed or delayed due to the depressed results.There are no reports of adverse health consequences due to the depressed advia centaur xp ft4 results.
 
Manufacturer Narrative
On 01/19/2017 additional information: recall number: 1219913-01/10/2017-002-r.
 
Manufacturer Narrative
Siemens filed the initial mdr 1219913-2016-00230 on november 18, 2016.On 01/05/2017 additional information: siemens has performed an internal investigation and has confirmed a negative bias for the advia centaur ft4 when used with calibrator a kit lots ending in 90 on the advia centaur, advia centaur xp and advia centaur xpt systems.In addition, siemens healthcare diagnostics confirmed the potential for calibration failures due to above limit calibrator rlu %cvs when using calibrator a kit lots ending in 90 with the ft4 assay.Siemens issued ufsn cc 17-04.A.Ous (january 2017) and umdr cc 17-04.A.Us (january 2017) informing the customer of ft4 assay negative bias observed with calibrator a kit lots ending in 90.Instructions on actions to be taken are provided in the customer communication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP FT4
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC
333 coney street
east walpole MA 02032
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
5086604381
MDR Report Key6113203
MDR Text Key60416252
Report Number1219913-2016-00230
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K132249
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Remedial Action Inspection
Type of Report Initial,Followup,Followup
Report Date 01/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/04/2017
Device Model NumberN/A
Device Catalogue Number06490106
Device Lot Number113072
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2016
Initial Date FDA Received11/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/13/2017
01/26/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/03/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1219913-01/10/2017-002-R
Patient Sequence Number1
-
-