Model Number P3700B000002 |
Device Problem
Self-Activation or Keying (1557)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/20/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The hill-rom technician has not yet evaluated the bed.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance (pm) along with a quarterly battery check and testing for (b)(4).Pm and testing not only meet (b)(4) requirements but will help ensure a long, operative life for the bed.Pm will minimize downtime due to excessive wear.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
|
|
Event Description
|
Hill-rom received a report from the account stating the bed hilow would self run.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
|
Manufacturer Narrative
|
The hill-rom technician found the shoulder bolt that holds the trend dampner on the right side was not screwed into the frame of the bed.The technician tightened the shoulder bolt back into the frame of the bed.Based on this information, no further action is required.
|
|
Event Description
|
Hill-rom received a report from the account stating the bed hilow would self run.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
|
|