• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INS SIZE 4 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INS SIZE 4 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531G409
Device Problems Device Operates Differently Than Expected (2913); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2016
Event Type  malfunction  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Additional information has been requested and if received, will be provided in the supplemental report upon completion of the investigation.
 
Event Description
The surgeon proceeded to insert a 9mm cs insert, like he always has.It had a small space on the lateral side of the tibia and poly.The surgeon tried twice to impact it with no luck.He asked for another cs 9mm poly and proceeded to impact the poly with the same result.They tried a cr 9mm poly and it settled correctly.
 
Manufacturer Narrative
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as the device was not returned for evaluation.-medical records received and evaluation: not performed as no medical records were provided.-device history review: indicates devices were manufactured and accepted into final stock with no reported discrepancies.-complaint history review: chr review for the reported lot confirmed that there are no other similar events reported.Conclusions: the exact cause of the event could not be determined because the devices were not returned for evaluation and no medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.No further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If additional information becomes available, this investigation will be reopened.
 
Event Description
The surgeon proceeded to insert a 9mm cs insert, like he always has.It had a small space on the lateral side of the tibia and poly.The surgeon tried twice to impact it with no luck.He asked for another cs 9mm poly and proceeded to impact the poly with the same result.They tried a cr 9mm poly and it settled correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INS SIZE 4 9MM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6115647
MDR Text Key60557246
Report Number0002249697-2016-03665
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number5531G409
Device Lot NumberLFD206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/23/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
-
-