Catalog Number 227425500 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/18/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
Drill bits broke approximately 10 - 15 mm distal of the tip while drilling from screws.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Examination of the returned drills confirmed the complaint; the majority of the fluted tips broke off.A complaint database search has identified a trend for this failure within the (b)(4) quickset drill family.Previous investigations have determined that use error is the likely root cause.As a result of trending health hazard evaluation, (b)(4) was conducted.The investigation did not establish the need for corrective action.No evidence was found of product or design error as a contributing factor.Complaints will be monitored under (b)(4) post market surveillance.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Search Alerts/Recalls
|