Model Number 4FC12
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2016
Event Type  malfunction  
Event Description
It was reported that during a cryoablation procedure, the sheath had bled back from the insertion port.The sheath was replaced.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files for the date of the reported event were returned and analyzed.The data files showed two system notices were received, both unrelated to the reported issue of air ingress due to the sheath.The data files also show seven injections were performed with one balloon catheter and nine injections were performed with another balloon catheter for the date of the event.The physical sheath device analysis is pending.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Visual inspection of sheath showed that the stopcock was intact with no apparent issues.Air aspiration was reproduced when a test catheter was introduced through the sheath.A dissection showed that the hemostatic valve was leaking and torn.In conclusion, the sheath failed the returned product inspection due to a leaking hemostatic valve.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire, qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire, qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6121422
MDR Text Key60687500
Report Number3002648230-2016-00498
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/12/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number67935
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-