Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Angina (1710); Stenosis (2263)
Event Date 09/22/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The three additional absorb scaffolds referenced are being filed under separate manufacturer report numbers.The absorb device is currently not commercially available in the u.S; however, it is similar to a device sold in the us.The device was not returned for evaluation.A review of the lot history record and complaint history could not be conducted because the part and lot numbers were not provided.The reported patient effects of angina and restenosis, as listed in the absorb instructions for use (ifu), are known adverse events associated with the use of a coronary scaffold in native coronary arteries.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.
 
Event Description
It was reported that index procedure on (b)(6) 2015 was performed at (b)(6) hospital on a patient with coronary artery disease.Four absorb scaffolds were implanted in the left anterior descending (lad) and left circumflex (lcx) vessels.On (b)(6) 2016, the patient experienced chest pain and had a re-intervention at (b)(6) hospital.Intravascular ultrasound (ivus) showed the lad scaffold(s) undersized with heavy plaque.The lcx scaffolds were was totally occluded with restenosis and the lad scaffolds had 80% restenosis.Three drug eluting stents were implanted for treatment with a good final patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6122583
MDR Text Key60743754
Report Number2024168-2016-08263
Device Sequence Number1
Product Code PNY
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2016
Initial Date FDA Received11/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCAFFOLD: ABSORB (X3)
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age45 YR