• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Lot Number E342-KIT
Device Problems Break (1069); Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/27/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review of kit lot e342 was conducted.There were no nonconformances associated with this lot.The lot met release requirements.The uvadex lot number was not provided as it was not administered.However, a review of all uvadex lots manufactured since january 2013 was performed.No trends or nonconformances related to the complaint were noted.Trends were reviewed for complaint categories, alarm #7: blood leak (centrifuge chamber), centrifuge bowl leak/break, and drive tube leak/break.No trends were detected for these complaint categories.An issue review was opened for complaint category, centrifuge bowl leak/break.Service order report, #(b)(4), feedback: the service technician cleaned the instrument and replaced the two drive tube retainer clips and also the centrifuge leak detector strip.The system checkout procedure was then successfully performed.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).
 
Event Description
The customer called to report an alarm #7: blood leak (centrifuge chamber) alarm.The customer stated that a combination centrifuge bowl leak/break and drive tube leak/break occurred at the start of the treatment.The customer reported that the treatment was aborted with no blood/products returned to the patient.The customer stated that the patient was in stable condition.The customer reported that there were no alarms prior to the break.Service was requested.The kit was not returned for investigation as it had already been discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERAKOS CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX
Manufacturer (Section D)
THERAKOS, INC
hampton NJ
Manufacturer (Section G)
THERAKOS, INC.
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
53 frontage road
hampton, NJ 08827
MDR Report Key6124695
MDR Text Key61100723
Report Number2523595-2016-00266
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date08/01/2018
Device Lot NumberE342-KIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient Weight93
-
-