MAUDE Adverse Event Report: NIHON KOHDEN TOMIOKA CORPORATION GF-210RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN TOMIOKA CORPORATION GF-210RA; MULTI-GAS UNIT Back to Search Results
Model Number GF-210RA
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reports that the carbon dioxide readings on the multi gas unit works intermittently.No patient harm was reported.The unit was cleaned and evaluated.The reported problem could not be duplicated through testing, trouble shooting, and extended operation of the device.Review of the device history indicates no previous cannot duplicate status.The unit was tested per the operator's manual and the results were recorded on the maintenance check sheet.The unit operates to manufacturer's specification.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reports that the carbon dioxide readings on the multi gas unit works intermittently.No patient harm was reported.Customer would like to send unit in for evaluation and repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GF-210RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: susan shadley
tomioka city, japan 370-2 314
JA  370-2314
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: susan shadley
tomioka city, japan 370-2 314
JA   370-2314
Manufacturer Contact
susan shadley
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8-560
JA   161-8560
2687133
MDR Report Key6127133
MDR Text Key60869877
Report Number8030229-2016-00544
Device Sequence Number1
Product Code CCK
UDI-Device Identifier04931921106891
UDI-Public04931921106891
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110594
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/24/2016,10/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-210RA
Device Catalogue NumberGF-210RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/24/2016
Distributor Facility Aware Date10/26/2016
Device Age19 MO
Event Location Hospital
Date Report to Manufacturer11/24/2016
Date Manufacturer Received11/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-