Model Number 382400 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 11/28/2016.An investigation is currently underway.Upon completion, a full report shall be provided.
|
|
Event Description
|
It was reported to covidien on (b)(6) 2016 that the customer experienced an issue with an enteral feeding pump.During triage, it was found that the unit would power on/off intermittently.
|
|
Manufacturer Narrative
|
The unit was triaged and the reported condition was confirmed.The power cycle issue is due to a faulty membrane overlay.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|