The actual sample involved in the reported incident was not returned for evaluation.There was no additional information, pictures or videos received.Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation.A device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhrs are reviewed for accuracy prior to product release.If the sample is returned in the future, this complaint will be re-opened for further investigation.No trends or triggers have been found, therefore, a corrective or preventive action is not deemed necessary at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for (b)(4) material, 100% in process visual inspection and visual acceptance sampling are performed in the plant) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
|