• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KIT 11.5FX19.5 CE MAHURKAR; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KIT 11.5FX19.5 CE MAHURKAR; DIALYSIS CATHETER Back to Search Results
Model Number 8813793013
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently under way; upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer had an issue with a dialysis catheter.The customer reports during the surgical process the guide wire did not pass through the catheter.Another catheter was used to conclude the procedure.The patient did not have any injuries.
 
Manufacturer Narrative
The actual sample involved in the reported incident was not returned for evaluation.There was no additional information, pictures or videos received.Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation.A device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhrs are reviewed for accuracy prior to product release.If the sample is returned in the future, this complaint will be re-opened for further investigation.No trends or triggers have been found, therefore, a corrective or preventive action is not deemed necessary at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for (b)(4) material, 100% in process visual inspection and visual acceptance sampling are performed in the plant) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT 11.5FX19.5 CE MAHURKAR
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS 
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS  
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6128616
MDR Text Key61204676
Report Number3009211636-2016-00473
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeMX
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service Personnel
Type of Report Initial,Followup
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8813793013
Device Catalogue Number8813793013
Device Lot Number1616100147
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-