• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701028718
Device Problem Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The technician observed the system and troubleshot to power supply.The technician rewired the 20 amp breaker connections and replaced the 20 amp breaker as a precaution.Unit passed all functional and safety tests per factory specifications, returned to customer and cleared for clinical use.A supplemental medwatch will be submitted if new information has been received.
 
Event Description
It was reported that the hcu 30 shut down during use.The incident occurred during patient treatment and the device was exchanged for another.There were no negative consequences to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6131440
MDR Text Key61401974
Report Number8010762-2016-00692
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701028718
Device Catalogue NumberMCP00939001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2016
Initial Date FDA Received11/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-