MAQUET CARDIOPULMONARY AG MAQUET HCU30 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701034371
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A maquet field service technician was on site and investigated the unit.The technician was unable to duplicate the problem.The technician started with a full block of ice and let the unit run for 2 hours at 37 degrees.The technician checked to ensure the compressor comes on.The unit passed all functional and safety tests as per factory specifications.A supplemental medwatch will be submitted if additional information becomes available.
 
Event Description
Customer stated that the compressor would not turn on to make ice when maintaining a setpoint of 37 degrees on both cardioplegia and main side.Additional information: the incident occurred during patient treatment, there were no negative consequences for the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU30 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6131474
MDR Text Key61401879
Report Number8010762-2016-00693
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701034371
Device Catalogue Number701034371
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-