BIOMET ORTHOPEDICS STRAIGHT MONOBLOCK ACETABULAR SHELL INSERTER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 01/01/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.Event date - 2016.There are warnings in the package insert that state that this type of event can occur: under care and handling of instruments, it states, "surgical instruments and instrument cases are susceptible to damage for a variety of reasons including prolonged use, misuse, rough or improper handling.".
 
Event Description
It was reported that the inserter fractured / chipped.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Failure mode was instrument fracture with the fracture pattern of the threads being consistent to that of bending overload.Inspection of the device showed evidence of heavy use and strike marks were evident at the bottom of the handle as if the device had been attempted to be hammered backwards by the handle.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT MONOBLOCK ACETABULAR SHELL INSERTER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6134619
MDR Text Key61195712
Report Number0001825034-2016-04913
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number636740
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A